Alco-Screen/Alco-Screen02™ digital alcohol test
Buy
Alco-Screen/Alco-Screen02 digital alcohol test
Procedure
- Abstain from placing anything
in the mouth for fifteen (15) minutes prior to beginning the test. This
includes non-alcoholic drinks, tobacco products, coffee, breath mints, food,
etc.
- Open the foil package
and remove the test strip. Observe the reactive pad on the end of the test
strip. The pad should be a light cream color. A test strip with a reagent
pad which is dark tan in color or otherwise discolored must be discarded.
- Saturate the reactive
pad with saliva from mouth or sputum cup. Immediately start timer.
- At two (2) minutes
observe the color change (if any) in the reactive pad. A color change of
green or blue indicates the presence of alcohol and a positive result.
Results obtained after more than 2 minutes and 30 seconds may be erroneous.
- Estimate the
approximate blood alcohol concentration by comparing the color of the
reagent pad with the color chart appearing on the test package.
Results
A ALCO-SCREENTM/ALCO-SCREEN02digital test produces a color change in the presence of saliva alcohol ranging from a light
green-gray color at 0.02% blood alcohol concentration to a dark blue-gray color
near 0.30% blood alcohol concentration. Color blocks are provided within this
range to allow an approximation of blood alcohol concentration to be made. ALCO-SCREENTM/ALCO-SCREEN02may produce colors that appear to be between adjacent color blocks.
ALCO-SCREENTM/ALCO-SCREEN02digital test are very sensitive to the presence of the drug alcohol. A green color that is lighter than
the 0.02% color block should be interpreted as being positive to the presence of
alcohol in saliva but less than 0.02% blood alcohol.